Abstract

The relative standard deviation (RSD) of measurements in high-performance liquid chromatography with electrochemical detection (HPLC–ECD) was predicted by a chemometric tool based on the 1/ f fluctuation model which is made up of white noise and a Markov process, called the Function of Mutual Information (FUMI) theory. FUMI theory can provide a precise and reliable detection limit from a single measurement of noise and signal in HPLC–ECD. To obtain RSD ( n=5) for determination of (−)-epicatechin at five concentrations required 12.5 h, while the predicted RSD by FUMI theory required only 0.5 h (one measurement). Moreover, to trace the source of instrumental noise, power spectra of chromatographic baseline were used. Selection of a suitable apparatus in HPLC–ECD system, acquisition of RSD, and detection limits for determination of catechins by HPLC–ECD were simply and easily made by this chemometric tool within a very short time. The use of the FUMI theory for the prediction of measuring precision was more efficient and the optimization was less time-consuming to be suited for determination.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call