Abstract

A chemometric tool based on the Function of Mutual Information (FUMI) theory can provide a relative standard deviation (RSD) without repetitive measurements in high-performance liquid chromatography with electrochemical detection (HPLC-ECD). Two parameters: precision (= information content) phi and efficiency (= information content/time) theta, which were calculated from predicted RSD based on the FUMI theory, were used to optimize HPLC-ECD conditions, such as applied potential, flow rate, column length, and size of ODS porous packing. We selected catechins as analytes, and found that the most optimum applied potential and flow rate were +600 mV vs. SCE and 0.9 mL/min, respectively, because they gave the largest phi and theta values. Buffer concentration in mobile phase is less effective for giving large phi and theta values. Since the FUMI theory makes it possible to predict RSD without repetitive measurements, the present method saves considerable amounts of chemicals and experimental time, and was found to be useful for the optimization of experimental conditions for determination by HPLC-ECD.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.