Abstract

AbstractFor the performance assessment of an electrochemical flow cell using a boron‐doped diamond (BDD) electrode in high‐performance liquid chromatography with electrochemical detection (HPLC‐ECD) to determine cholesterol, the standard deviation (SD) of area measurements in chromatographic baseline noise (AMCBN) was evaluated based on the function of mutual information (FUMI) theory without repetitive chromatographic measurements. After the selection of suitable components for the electrochemical flow cell in HPLC‐ECD, the detection limit of cholesterol was 0.04 pmol at a detection potential of +2.2 V vs. an Ag/AgCl electrode. The present HPLC‐ECD was applied to determine cholesterol in various meats.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.