Abstract
When generating automatic sleep reports with mobile sleep monitoring devices, it is crucial to have a good grasp of the reliability of the result. In this paper, we feed features derived from the output of a sleep scoring algorithm to a 'regression ensemble' to estimate the quality of the automatic sleep scoring. We compare this estimate to the actual quality, calculated using a manual scoring of a concurrent polysomnography recording. We find that it is generally possible to estimate the quality of a sleep scoring, but with some uncertainty ('root mean squared error' between estimated and true Cohen's kappa is 0.078). We expect that this method could be useful in situations with many scored nights from the same subject, where an overall picture of scoring quality is needed, but where uncertainty on single nights is less of an issue.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.