Abstract

We have attempted to quantify the influence of clinical, radiological and prosthetic design factors upon flexion following knee replacement. Our study examined the outcome following 101 knee replacements performed in two prospective randomized trials using similar cruciate retaining implants. Multivariate analyses, after adjusting for age, sex, diagnosis and the type of prosthesis revealed that the only significant correlates for range of movement at 12 months were the difference in posterior condylar offset ratio (p < 0.001), tibial slope (p < 0.001) and preoperative range of movement (p = 0.025). We found a moderate correlation between 12-month range of movement and posterior tibial slope (R = 0.58) and the difference of post femoral condylar offset (i.e, post-operative minus preoperative posterior condylar offset, R = 0.65). Posterior condylar offset had the greatest impact upon final range of movement highlighting this as an important consideration for the operating surgeon at pre-operative templating when choosing both the design and size of the femoral component.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.