Abstract

The aim of this study was to examine the lifespan or number of cycles to failure of tapered rotary nickel-titanium (Ni-Ti) endodontic instruments. Simulated root canals with different curvatures were used to determine a relation between canal curvature and instrument lifespan. Using a novel mathematical model for the deformation of pseudoelastic Ni-Ti alloy, it was shown that maximum stress need not necessarily occur at the outer layer. On the basis of this observation, the Coffin-Manson relation was modified with parameters determined from this experiment. Results showed that the number of cycles to failure was influenced by the angle and radius of canal curvature and the size of instrument at the beginning of canal curvature. The resulting quantitative mathematical relation could be used to predict the lifespan of rotary Ni-Ti endodontic instruments under clinical conditions and thereby reduce the incidence of instrument failure in vivo.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.