Abstract

To investigate the bending properties of hybrid rotary nickel-titanium endodontic instruments in relation to their transformation behaviour. Four types of nickel-titanium rotary endodontic instruments with different cross-sectional shapes (triangular-based and rectangular-based) and different heat treatment conditions (super-elastic type and hybrid type with shape memory effect) were selected to investigate bending properties and phase transformation behaviour. Bending load of the instruments was measured in a cantilever-bending test at 37 degrees C with the maximum deflection of 3.0 mm. A commercial rotary instrument, ProFile (PF; Dentsply Maillefer, Ballaigues, Switzerland) was used as a reference for the bending test. Phase transformation temperatures were calculated from the diagrams obtained from differential scanning calorimetry. Data were analysed by anova and Scheffe's test. The bending load values of the hybrid type that had undergone additional heat treatment at the tip were significantly lower (P < 0.05) than those of the super-elastic type with no additional heat treatment. The bending load values of rectangular-based cross-sectional shaped instruments were significantly lower (P < 0.05) than those of triangular-based cross-sectional shaped instruments. Phase transformation temperatures (M(s) and A(f) points) of the hybrid type were significantly higher (P < 0.05) than the super-elastic type. The M(f) and A(s) points of the tip part were significantly higher (P < 0.05) than those of the whole part of the hybrid instrument. Additional heat treatment of hybrid nickel-titanium instruments may be effective in increasing the flexibility of nickel-titanium rotary instruments.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call