Abstract
A preconcentration method of lead in waters by adsorption on a tantalum wire was developed for electrothermal atomization atomic absorption spectrometry with a tungsten tube atomizer. After the preconcentration, the tantalum wire was directly inserted into the tungsten tube atomizer. In the preconcentration (adsorption) process for lead, the optimal immersing time was 90 s and the best pH was 4. Under the optimal conditions, the detection limit for lead by the tantalum wire preconcentration method was 6.0 pg mL − 1 (3S/N) and the relative standard deviation was 6.1%. The influences of large amounts of concomitants on the preconcentration of lead were evaluated. Even though 10 3 to 10 4-fold excess of matrix elements existed in aqueous solution, the lead absorption signal was not significantly affected by the matrix elements. The method with preconcentration on a tantalum wire was applied to the determination of lead in river waters and proved to be sensitive, simple, and convenient. Because this preconcentration method can be utilized in the in-situ treatment of trace lead in environmental water samples, it was unnecessary to carry the water samples to the analytical work place. The present technique was shown to be useful for the determination of lead in environmental water samples at 0.1−1 μg L − 1 .
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.