Abstract

Relative intensities of gamma rays were measured in Hiroshima and Nagoya in the region of 122–356 keV. Two Ge(Li) detectors were calibrated with four kinds of standard sources and a cascade gamma-ray source. The disintegration rates of the standard sources were determined by means of the coincidence method. Relative intensities of 75Se and 133Ba were obtained with estimated errors less than 1% for strong gamma rays. Gamma-ray intensities per decay were also extracted.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.