Abstract

The purpose of this cross-sectional study was to determine the precision and accuracy of the measurement of finger motion with a depth camera. Fifty-five healthy adult hands were included. Measurements were done with a depth camera and compared with traditional manual goniometer measurements. Repeated measuring showed that the overall repeatability and reproducibility of extension measured with the depth camera were within 3° and 4° and that of flexion were within 13° and 14°. Compared with traditional manual goniometry, biases of extension of all finger joints and flexion of metacarpophalangeal joints were less than 5°, and the average bias of flexion of proximal and distal interphalangeal joints was 29°. We conclude that the measurement of finger extension and flexion of the metacarpophalangeal joints with a depth camera was reliable, but improvement is required in the precision and accuracy of interphalangeal joint flexion.