Abstract

A practical technique to correct the dead time effect in the variance-to-mean ratio measurement is proposed. A correction formula used in the technique is derived based on a simple mechanism of count loss due to a non-extending dead time. The technique was verified with a Monte Carlo simulation and an experiment in Deuterium Critical Assembly. It was confirmed that the formula can make a reasonable correction to the case that the count loss is as large as 50%. This correction technique is extremely simple and can make the correction only with data obtained in the original measurement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.