Abstract

With the development of fast spin echo (FSE) MRI techniques, T2-weighted images of the brain may be obtained much more quickly than when using conventional spin echo techniques (CSE), because made the individual echoes on the FSE pulse sequence are phase encoded, allowing acquisition of the same spatial information as in CSE with less excitations. The pulse sequence parameters (echo train length, bandwidth, echo spacing) are discussed. Images were obtained on four volunteers using both CSE and FSE while varying repetition time, echo time and matrix. Comparison for signal intensity gray-white differentiation, fat and CSE signal, arifacts and vascular resolution showed that FSE images comparable in quality to those of CSE can be obtained in less than half the time. A practical choice of FSE parameters is recommended for clinical use. However, artifacts, possibly related to CSF and vascular pulsation, of which the radiologist should be aware, were identified on the FSE images.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.