This paper presents an optimized T-type resonant photoacoustic (PA) cell for methane (CH4) gas detection. The noise transmission coefficients and PA field distributions of the T-type resonant PA cell have been evaluated using the finite element method and thermoviscous acoustic theory. The optimized T-type resonant PA cell, together with a near-infrared (NIR) distributed feedback (DFB) laser source, a high-speed spectrometer and a fiber-optic acoustic sensor constitutes a PAS system for CH4 detection. The sensitivity is measured to be 1.8 pm/ppm and a minimum detectable limit (MDL) of 9 parts per billion (ppb) can be achieved with an averaging time of 500 s. The optimized T-type longitudinal resonant PA cell features of high PA cell constant, fast response time and simple manufacturing process.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.