Abstract

Potentiometric detection employing coated-wire electrodes was applied to the determination of organic acids in liquid chromatography (LC). Poly(vinyl chloride)-based liquid membranes, incorporating lipophilic macrocyclic hexaamines as neutral ionophores were used as electrode coatings. The selectivity and sensitivity of the macrocycle-based electrodes were found to be superior to an electrode based on a lipophilic anion exchanger (a quaternary ammonium salt). Sensitive detection was obtained for the di- and tricarboxylic acids tartaric, malonic, malic, citric, fumaric, succinic, pyruvic, 2-oxoglutaric and maleic acids after separation in reversed-phase LC. Detection limits (signal/4 σ noise=3) of 6 pmol for malonic acid and 2 pmol for maleic acid were attained. The detection was explained using a molecular recognition model. The hexaamine-based potentiometric electrodes had a 1-s response time at 1 ml min −1 flow-rates. They were stable for at least 4 months, with an intra-electrode variation of 3.2% ( n=5).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.