Abstract
A chronoamperometric method for obtaining experimental values of diffusion coefficients with high trueness and precision was described. In this method, the chronoamperograms were analyzed by using the most precise theoretical equation available. All of the variables that appear in the equation had been absolutely determined with instruments calibrated traceably. Experiments were carried out so conscientiously that the demands of the theoretical equation would be realized as well as possible. The obtained chronoamperogram was fitted to the theoretical equation by two adjustable parameters : the diffusion coefficient and the concentration of the diffusing spices. Each chronoamperogram was inspected concerning how it fitted the theoretical equation over the observed time range and how the obtained concentration agreed with the actually known one. By this inspection the trueness of the value of the diffusion coefficient was assessed. For an example of showing the importance of the trueness of the experimental values, the diffusion coefficients of fullerene in references were shown along with the values obtained by this method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.