Abstract
A large range of laser-matter applications employ ultrashort pulses and high laser intensity. Such processes can lead to unrequired X-ray generation, which represents a hazardous radiation factor even for common laboratory research-grade laser systems. We present here an analysis of the radiation dose rate and X-ray spectrum emitted during ablation of a rotating copper cylinder with respect to several laser parameters. The results show that focused sub-picosecond pulses with intensity above 1013 W/cm2 can exceed the annual irradiation limit even in one hour, requiring appropriate shielding for the safety of the researchers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.