Abstract
Benign, giant cell tumors are often treated by intralesional excision and reconstruction with polymethylmethacrylate (PMMA) bone cement. The exothermic reaction of the in-situ polymerizing PMMA is believed to beneficially kill remaining tumor cells. However, at issue is the extent of this necrotic effect into the surrounding normal bone and the adjacent articular cartilage. Finite element analysis (ABAQUS 6.4-1) was used to determine the extent of possible thermal necrosis around prismatically shaped, PMMA implants (8–24 cc in volume), placed into a peripheral, sagittally symmetric, metaphyseal defect in the proximal tibia. Temperature/exposure time conditions indicating necrotic potential during the exotherm of the polymerizing bone cement were found in regions of the cancellous bone within 3 mm of the superior surface of the PMMA implant. If less than 3 mm of cancellous bone existed between the PMMA implant and the subchondral bone layer, regions of the subchondral bone were also exposed to thermally necrotic conditions. However, as long as there were at least 2 mm of uniform subchondral bone above the PMMA implant, the necrotic regions did not extend into the overlying articular cartilage. This was the case even when the PMMA was in direct contact with the subchondral bone. If the subchondral bone is not of sufficient thickness, or is not continuous, then care should be taken to protect the articular cartilage from thermal damage as a result of the reconstruction of the tumor cavity with PMMA bone cement.