Abstract

Posterior-stabilized antibiotic cement articulating spacers (PS spacers) reduce spacer mechanical complications in prosthetic knee infections (PKIs); however, joint dislocation after femoral cam fracture has been reported. We hypothesized that the rate of post-cam mechanical complications is lower in PS spacers with an endoskeleton-reinforced cam. A retrospective study of PKIs using PS spacers with or without a Kirschner wire-reinforced cam (K-PS or nK-PS spacers, respectively) was conducted between 2015 and 2019. The rates of post-cam mechanical complications and reoperation, as well as risk factors for post or cam failure, were analyzed. The cohort included 118 nK-PS and 49 K-PS spacers. All patients were followed up for 2 years. The rate of joint subluxation/dislocation after femoral cam fracture was lower in K-PS (0%) than in nK-PS spacers (17.8%; P= .002). The reoperation rate for spacer mechanical complications was lower in K-PS (0%) than in nK-PS spacers (11.9%; P= .008). The identified risk factors for femoral cam fractures were body mass index ≥25 kg/m2, femoral spacer size ≤2, and surgical volume ≤12 resection arthroplasties per year. This preliminary study highlights that K-PS spacers have a lower rate of post-cam mechanical complications than nK-PS spacers. We recommend the use of PS spacers with endoskeleton-reinforced cam when treating PKIs performed by surgeons with lower surgical volumes, especially in patients with higher body mass index and smaller femoral spacer sizes.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call