Abstract

Twenty-two retrieved femoral knee components were identified with posterior condyle surface damage on average at 99° flexion (range, 43°-135° flexion). Titanium alloy material transfer and abrasive surface damage were evident on cobalt-chromium alloy femoral components that were in contact with titanium alloy tibial trays. Surface damage on the retrieved Oxinium femoral components (Smith and Nephew, Inc, Memphis, Tenn) that were in contact with titanium alloy tibial trays showed gouging, associated with the removal and cracking of the oxide and exposure of the zirconium-niobium alloy substrate. Cobalt-chromium alloy femoral components that were in contact with cobalt-chromium alloy tibial trays showed abrasive wear. Contact between the femoral component and tibial tray should be avoided to prevent surface damage to the femoral condyles, which could potentially accelerate polyethylene wear in vivo.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.