Abstract

In projection x‐ray imaging, in the absence of a grid scattered x‐ray photons from the patient out‐number primary photons at the image receptor by as much as 10:1. Scattered radiation can significantly degrade projection contrast and signal‐to‐noise ratio in conventional projection imaging, but can also be used as a source of additional information about the patient or object examined. In principle, a raster scan by a primary pencil beam and two‐dimensional scatter field detector could be used to acquire the scatter information. Simultaneous measurements would reduce acquisition time. Here we report on our first investigations of using a fan beam. We used a monoenergetic radiation beam at the BioMedical Imaging and Therapy (BMIT) facility at the Canadian Light Source (CLS) synchrotron (Saskatoon, Saskatchewan). X rays of approximately 40 keV produced by the 05B1–1 bend magnet beamline were used with Computed Radiography plates as the detector. The fan beam was 1.1 mm thick by 10.6 cm wide. The circularly‐symmetric diffraction patterns of water, plastics, and porcine tissue samples were sampled outside of the fan plane. Signal processing based on cross‐correlation of the scatter field with arc patterns was investigated as a means to provide approximate localization of materials in the fan plane. A more sophisticated signal‐processing approach and/or collimation strategy will likely be necessary.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.