Abstract

A heavy ion gas detector system consisting of a Bragg-curve spectroscopy ionization chamber for particle identification and a multiwire proportional chamber as position sensitive fast trigger device is described. The Bragg IC has been tested with several beams up to Z = 36 to investigate some aspects of the BCS method. Results are reported on energy resolution and linearity, Z resolving power and mass sensitivity. The energy resolution is well below 1%. The Bragg-peak amplitude is fairly independent of the energy in a wide energy range and single elements are identified up to Z = 38 with a resolving power Z/ ΔZ − 50-80. Isotope identification by range measurement is limited by the straggling in the ionization process and the mass resolving power is M/ ΔM ∼ 20–26 for S and Si isotopes. The MWPC allows subnanosecond time resolution and position identification along the in-plane coordinate within ±0.5 mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.