Abstract

Introduction: Volumetric Arc Radiotherapy (VMAT) is an advanced technique. Calculations of VMATplans are not so accurate even with State-of-Art dose calculation algorithms due to their complexity.Hence pre-treatment patient specific Quality Assurance (QA) of each VMAT plan is required. In thepresent study Electronic Portal Imaging Device (EPID) based portal dosimetry system was used forpre-treatment patient specific QA. Material and Methods: A total of 50 patients were chosen inthis study. Verification plans of each patient were calculated for portal dosimetry then executed onthe EPID system to measure the spatial distribution of radiation dose. Calculated and measured dosedistribution were compared to evaluate Gamma Index (GI) passing criteria of Dose Difference (DD)of 3% and Distance–to-Agreement (DTA) of 3mm, Area Gamma (γ% ≤1) >95%, Average Gamma(gAve) <0.5% and Maximum Gamma (gMax) <3.5%. Results: The mean values of Area Gamma (γ%≤1) were observed to be varied from 99.14±0.23% to 99.87±0.18%. The Mean Values of AverageGamma (gAve) are found to vary from 0.19±0.05% to 0.15±0.04% and the mean values ofMaximum Gamma (gMax) found to be varied from 1.94±0.37% to 1.59±0.41%. All the plans werepassed the gamma index criteria with very good agreement. Thus the use of Portal Dosimetry forpre-treatment patient QA is found to be a very useful, quick, precise, efficient and effective pre-treatment patient specific QA tool for VMAT treatment. Conclusion: Portal Dosimetry can be utilizedfor routine use for patient specific quality assurance for Volumetric Arc Radiotherapy treatment.

Highlights

  • Volumetric Arc Radiotherapy (VMAT) is an advanced technique

  • Portal Dosimetry can be utilized for routine use for patient specific quality assurance for Volumetric Arc Radiotherapy treatment

  • The gamma index (GI) evaluation of the measured parameters are within the tolerance limit which dose distributions in the Portal dosimetry/Electronic Portal Imaging Device (EPID)

Read more

Summary

Introduction

Volumetric Arc Radiotherapy (VMAT) is an advanced technique. Calculations of VMAT plans are not so accurate even with State-of-Art dose calculation algorithms due to their complexity. In the present study Electronic Portal Imaging Device (EPID) based portal dosimetry system was used for pre-treatment patient specific QA. Conclusion: Portal Dosimetry can be utilized for routine use for patient specific quality assurance for Volumetric Arc Radiotherapy treatment. Calculations of the small or irregular fields which are frequently used for VMAT are not accurate even with the state-of-the-art dose calculation algorithm [8], patient specific Quality Assurance (QA) of every patient treated with VMAT techniques were performed before patient treatment [9,10,11]. In the present study EPID based portal dosimetry system was used as a pre treatment patient specific QA for VMAT plans treated at the Department of Radiotherapy, Regional Cancer Centre, Pt. J.N.M. Medical College, Raipur

Methods
Results
Conclusion