Abstract

A novel sorbent, dihydroxylated polymethylmethacrylate (DHPMM), coated on hollow-fiber membrane, is used for the polymer-coated hollow-fiber microextraction of trace amounts of natural and synthetic estrogens, such as diethylstilbestrol, estrone, 17β-estradiol and 17α-ethynylestradiol, in aqueous samples. In this procedure, estrogens were extracted using the functionalized polar DHPMM polymer with derivatization using N-methyl- N-(trimethylsilyl)trifluoroacetamide followed by gas chromatography–mass spectrometric analysis. The detection limits for estrogens in aqueous sample were between 0.03 and 0.8 ng l −1 and the calibration curves were linear over the concentration range 0.05–10 μg l −1 and had correlation coefficients of >0.994. The relative standard deviations (RSDs) were <15% ( n = 3). This simple, accurate, sensitive and selective analytical method is applicable to the determination of trace amounts of estrogens in reservoir and potable water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.