Abstract

A novel and simple method is presented for the preconcentration and determination of mercury (Hg) from natural waters through its extraction into a polymer inclusion membrane (PIM) containing the task-specific ionic liquid trioctylmethylammonium thiosalicylate (TOMATS) followed by Energy Dispersive X-ray fluorescence (EDXRF) analysis. The determination was made directly on the membrane without any treatment or elution step, and due to the characteristics of the PIM no matrix or thickness corrections were required in EDXRF analysis. Under the best extracting and EDXRF operating conditions, a Hg limit of detection of 0.2 μg Hg L-1 was obtained. Moreover, no water matrix effect was observed when Hg was extracted from different types of water such as river, seawater, groundwater, and tap water, showing this extraction system as a global solution when dealing with natural waters. Interestingly, this Hg collected in the PIM has shown to be stable for at least 6 months without the use of any preservative. This fact is of prime importance taking into account the usual stability problems of Hg during sample storage.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.