Abstract

The Standard Thermal Neutron Flux Unit, TNF2, in (LNMRI/IRD)(1, 2), was built for neutron detector and survey meter calibrations. The facility's fluence is achieved by moderation of four 241Am-Be with a graphite core and paraffin/graphite blocks surrounding it. Due to the small channel dimensions, it is impossible to calibrate personal dosemeters and survey meters for thermal neutrons. A polyethylene filter construction was carried out to allow the external irradiation of personal dosemeters and neutron survey meters. The polyethylene filter was constructed with 29 stacked discs with diameters ranging from 5 to 34cm. Different thicknesses were simulated to provide the desired effect. This new irradiation configuration was also experimentally tested and compared with simulation results with MCNPX(3).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.