Abstract

Poly(n-butyl acrylate) membranes for potentiometric ion-selective electrodes were developed and studied on example of lead-selective sensors. A novel approach resulting in Nernstian responses of tested sensor was proposed. Introduction of 5% (w/w) hydroxyethyl methacrylate into n-butyl acrylate moiety resulted in significant improvement of sensor analytical parameters. For the latter membrane material linear responses were obtained within lead activities range from 10 −2 to 10 −9 mol/dm 3, while for poly(n-butyl acrylate) based membranes pretreated in the same manner super-Nernstian behavior was obtained in a parallel experiment. Electrochemical impedance spectroscopy studies did not reveal significant differences between these two membranes, also similar lead ions diffusion coefficients were determined using inductively coupled plasma mass spectrometry with laser ablation.The difference between two kinds of membranes was found to concern higher Pb(II) ions contents in the surface part of the membrane with hydroxyethyl methacrylate, resulting in balanced Pb ions fluxes from/to the membrane.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.