Abstract

The grating interferometer in the Littrow configuration uses quarter wave plates (QWPs) to modulate the polarization in the measurement system to determine the autocollimation optical path. Fabrication errors and mounting errors of the QWPs lead to phase changes in the grating interferometer that generate measurement errors. As an alternative, we propose a grating interferometer that produces conical diffraction. Using the grating instead of QWPs to modulate the beam's polarization bypasses this source of error. A 45 mm range experiment was performed that yielded a repeated measurement error of 40 nm. Experiments show that the system has a simple structure and good repeatability and is capable of high-precision displacement measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.