Abstract

This paper presents the use of a point spectroscopy system to determine one’s transcutaneous bilirubin level using Modified Lambert Beer model and the developed fitting routine. This technique required a priori knowledge of extinction coefficient of bilirubin and hemoglobin components in the wavelength range of 440-500 nm for the prediction of the required parameter value. This work was conducted on different skin sites of six healthy Asians namely on the thenar region of the palm of their hand, back of the hand, posterior and anterior forearm. The obtained results revealed the lowest mean transcutaneous bilirubin concentration of 0.44±0.3 g/l predicted for palm site while the highest bilirubin level of 0.98±0.2 g/l was estimated for posterior forearm. These values were also compared with that presented in the literature. This study found considerably good consistency in the value predicted for different subjects especially at the thenar region of the palm. This work concluded that the proposed system and technique may be suitably served as an alternative means to noncontact and noninvasive measurement of one’s transcutaneous bilirubin level at palm site.