Abstract

BackgroundQuantification of macrosteatosis (MS) in the liver is important given that it has shown to directly correlate with adverse post-liver transplant (LT) outcomes. With advances in medical technology and an implicit understanding of pathology, noninvasive methods of quantitatively assessing MS are in various stages of development. Each of these methods is based on the physical principles of differences between a fat-laden hepatocyte and a normal one. MethodsIn this regard, after a proof-of-concept study on a prototype for a simple, real-time, handheld device using the principle of diffuse reflectance spectroscopy, this study presents an upgraded point-of-care (POC) device for the noninvasive assessment of hepatic MS in liver donors. ResultsThe device was validated on cohort of donor livers and showed a sensitivity (0.0021 V/% fat) and highly correlated (r = 0.9868, P < .0001) with gold-standard liver biopsy. Results showed that this upgraded POC device provides a reliable method for the noninvasive assessment of hepatic MS, which is crucial for selecting suitable donor livers for LT. ConclusionThe device has the potential to be an invaluable apparatus at the hands of the organ-retrieving surgeon. It is noninvasive, portable (handheld), and economic; provides real-time readings of the percentage of MS; and can be efficaciously handled by any member of the organ-retrieving team.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.