Abstract
BackgroundRespiratory variations in pulse oximetry plethysmographic waveform amplitude (ΔPOP) can predict fluid responsiveness in mechanically ventilated patients but cannot be easily assessed at the bedside. Pleth variability index (PVI) is a new algorithm allowing for automated and continuous monitoring of ΔPOP. We hypothesized that PVI can predict fluid responsiveness in mechanically ventilated patients under general anaesthesia. MethodsTwenty-five patients were studied after induction of general anaesthesia. Haemodynamic data [cardiac index (CI), respiratory variations in arterial pulse pressure (ΔPP), ΔPOP, and PVI] were recorded before and after volume expansion (500 ml of hetastarch 6%). Fluid responsiveness was defined as an increase in CI ≥15%. ResultsVolume expansion induced changes in CI [2.0 (sd 0.9) to 2.5 (1.2) litre min−1 m−2; P<0.01], ΔPOP [15 (7)% to 8 (3)%; P<0.01], and PVI [14 (7)% to 9 (3)%; P<0.01]. ΔPOP and PVI were higher in responders than in non-responders [19 (9)% vs 9 (4)% and 18 (6)% vs 8 (4)%, respectively; P<0.01 for both]. A PVI >14% before volume expansion discriminated between responders and non-responders with 81% sensitivity and 100% specificity. There was a significant relationship between PVI before volume expansion and change in CI after volume expansion (r=0.67; P<0.01). ConclusionsPVI, an automatic and continuous monitor of ΔPOP, can predict fluid responsiveness non-invasively in mechanically ventilated patients during general anaesthesia. This index has potential clinical applications.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have