Abstract
Calibration of an up-down type Mach probe is performed using a fast-flowing plasma produced by a magneto-plasma-dynamic arcjet. Mach probe data are compared with ion acoustic Mach numbers Mi, which are calculated using a plasma flow velocity Up and an ion temperature Ti measured by spectroscopy and electron temperature Te by Langmuir probe. The obtained data are also compared with Hutchinson’s PIC simulation results in an unmagnetized plasma and are in good agreement with each other. First attempts to measure plasma flow field using a Mach probe are performed at the open-end section in GAMMA10. It is found that Mi at r=0 is more than 2, namely a supersonic plasma flow is formed in the end-cell region.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.