Abstract
Background Clinical trials of measles vaccination administered as aerosol are planned with the aim of obtaining licensure. Measles antibody levels will be measured using the plaque reduction neutralization test (PRNT) to assess antibody responses as a surrogate marker of efficacy. Methods A working group examined laboratory protocols for measles PRNT in use at three reference centres and agreed to a standardised procedure, which was subsequently validated. Results Assay validation showed quantitative results varied approximately threefold both within and between assays. The lower limit of detection was approximately 20 milliInternational Units/mL. Conclusions A standardised laboratory protocol for measles PRNT was established and validated for use in clinical trials of aerosolized measles vaccines.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.