Abstract
BackgroundManaging inter-fractional anatomy changes is a challenging task in radiotherapy of pancreatic tumors, especially in scanned carbon-ion delivery. This treatment planning study aims to focus on clinically feasible solutions, such as the beam angle selection and margin design to increase the robustness against inter-fractional uncertainties.MethodsThis study included 10 patients with weekly 3D-CT imaging and physician-approved Clinical Target Volume (CTV). The study was directed to keep the CTV-coverage using six beam angle configurations in combination with different Internal Target Volume (ITV) concepts. These were: geometric-margin (symmetric 3 and 5 mm margin); range-equivalent margins with an isotropic HU replacement; and to evaluate the need of asymmetric margins the water-equivalent range path (WEPL) was determined per patient from the set of CTs.Plan optimization and forward dose calculation in each week-CT were performed with the research treatment planning system TRiP98 and the plan quality evaluated in terms of CTV coverage (V95CTV) and homogeneity dose (HCTV = D5-D95).ResultsThe beam geometry had a substantial impact on the target irradiation over the treatment course, with the single posterior or two beams showing the best average coverage of the CTV. The use of geometric margins for the more robust beam geometries showed acceptable results, with a V95CTV of (99.2 ± 1.2)% for the 5 mm-margin. For the non-robust configurations, due to substantial changes in the radiological depth, the use of this margin results in a V95CTV that might be below 80%, only showing improvement when the range changes are included.ConclusionsSelection of adequate beam configurations and treatment margins in ion-beam therapy of pancreatic tumors is of great importance. For a single posterior beam or two beam configurations, application of geometrical margins compensate for dose degradation induced by inter-fractional anatomy changes for the majority of the analyzed treatment fractions.
Highlights
Pancreatic cancer is still a disease without an effective treatment option usually with low survival rates and local control [1]
To extract the information of the margins direction and size we evaluated the variation of the center of mass (COM) and the Hausdorff distance (HD) between ITVWEPL and Clinical Target Volume (CTV) [16], which identifies the largest of all distances from one of the contours to the closest point on the other contour [17]
These values are related with the observed range uncertainties in the anterior abdomen region due to the inter-fractional variability of the anatomy
Summary
Pancreatic cancer is still a disease without an effective treatment option usually with low survival rates and local control [1]. The high sensitivity of the ion range to density changes in the beam-path induces dose underand overshoots. These variations can either result from inter-fractional anatomy changes, patient positioning or intra-fractional motion [3]. Managing inter-fractional anatomy changes is a challenging task in radiotherapy of pancreatic tumors, especially in scanned carbon-ion delivery. This treatment planning study aims to focus on clinically feasible solutions, such as the beam angle selection and margin design to increase the robustness against inter-fractional uncertainties
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.