Abstract

An arteriovenous fistula is artificially produced using a graft for hemodialysis in patients. In an arteriovenous graft (AVG), the angle of its arterial or venous anastomosis play an important role in producing flows inside blood vessels, through which a stenosis may occur. Most studies thus far have focused on CFD results. In this study, a PIV technique is used to analyze the hemodynamic characteristics at the arterial or venous anastomosis of an AVG having an angle of . For flow dynamic similarity, the Reynolds number is set to be the same for real and simulated flows. A PIV experiment is performed with a control valve in the arterial part. In conclusion, the recirculation flow appeared in the bifurcation area and the total blood velocity changed according to the extent of valve opening.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.