Abstract
For the routine determination of residues of growth promoters two important types of low-resolution gas chromatograph—mass spectrometer may be distinguished: the ultra-trace full-scan instrument [e.g., the ion trap mass spectrometer (ITS40)] and most other quadrupole apparatus using the selected ion monitoring (SIM) mode for detecting very small amounts (< 1–10 ng). In analysing biological extracts interference between matrix components, present at high concentrations, and analytes, present at low concentrations, should be avoided. In this investigation theoretical examples of pitfalls in SIM due to isotope interference ( 13C) from matrix components with the analyte were considered. These interferences may lead to false-positive and -negative results and false quantification.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.