Abstract

An isotope dilution mass spectrometric assay for the quantification of glutaric acid in urine and serum samples was developed. The performance of a quadrupole mass filter (QMF) gas chromatography–mass spectrometry (GC–MS) instrument, operated in the selected ion monitoring mode, and a quadrupole ion trap (QIT) GC/MS instrument, operated in the selected ion storage mode, was compared. Both instruments gave linear standard curves with glutaric acid concentrations between 0.19 and 3.8 μM. The average coefficients of correlation were 0.9998 and 0.9993 for the QMF and the QIT system, respectively. There was good agreement between the glutaric acid concentrations measured with the two instruments. The run-to-run precision was between 1.2 and 3.7% and between 6.2 and 8.6%, the average recovery of glutaric acid in urine and serum samples was 96 and 103% with the QMF and QIT instrument, respectively. We conclude that although the QMF has a slightly better performance, both instruments can be used to reliably measure glutaric acid concentrations from urine and serum patient samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.