Abstract

The ultrasonic surgical aspirator employs a vibrating metal tip to fragment tissue and then aspirates the debris through the hollow center of the tip. The mechanism of interaction has been stated to be poorly understood, most likely related to cavitation, possibly acting in concert with other mechanical actions. The role of stroke, suction, frequency, tissue type, and tip area have been examined with regard to tissue fragmentation rate. Suction is shown to make a significant contribution to the interaction. Photographic and acoustic data from experiments in water and on a range of fresh pig tissues are used to investigate the fragmentation effect. A model for the primary mechanism for tissue fragmentation is presented. This involves the horn-tip impact and other mechanical forces, operating in combination with hydrodynamic forces applied to the tissue on the forward stroke in each cycle. No evidence of cavitation in tissue was observed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.