New catalase- and peroxidase-type iron porphyrin biomimetic electrodes have been developed for determining ultralow concentrations of H2O2 and C2H5OH in aqueous solutions. Their physicochemical features have been studied. A mechanism of catalase and peroxidase reactions was suggested. Biomimetic electrodes did not lose their activity for a long time under the action of the oxidant, intermediates, and the final products of the decomposition of H2O2. Potentiometric biomimetic sensors of catalase and peroxidase types have been designed and studied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.