Abstract

In this paper the effect of treatment room and maze layout on the photoneutron and capture gamma dose equivalent in the maze was studied. MCNPX Monte Carlo (MC) code was used to simulate the Varian 2100 C/D Clinac 18 MV and four different room layouts. Two analytical methods, Wu-McGinley and McGinley, were used for dose calculations. The analytical methods overestimated the photoneutron dose (13-43 %) and gamma capture dose (16-95 %) comparing with the MC method at the maze entrance door. The results of MC method revealed that additional bend can cause a great reduction in photoneutron (5000 times) and capture gamma dose (50 times) in the maze entrance door.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.