Abstract

Pencil beam kernels describing scattered photon fluence behind homogeneous water slabs at various air gap distances were generated using the EGS Monte Carlo code. Photon scatter fluence was scored in separate bins based on the particle's history: singly scattered, multiply scattered, and bremsstrahlung and positron annihilation photons. Simultaneously, the mean energy and mean angle with respect to the incident photon pencil beam were tallied. Kernels were generated for incident photon pencil beams exhibiting monoenergetic spectra of 2.0 and 10.0 MeV, and polyenergetic spectra representative of 6 and 24 MV beams. Reciprocity was used to generate scatter fractions on the central axis for various field sizes, phantom thicknesses, and air gaps. The scatter kernels were further characterized by full width at half-maximum estimates. Modulation transfer functions were calculated, providing theoretical estimates of the limit of performance of portal imaging systems due to the intrinsic scattering of photon radiation through the patient.