Abstract

X and gamma rays continue to remain the main contributors to the dose to humans. As these photons of varying energies are encountered in various applications, the study of photon energy response of a dosemeter is an important aspect to ensure the accuracy in dose measurement. Responses of dosemeters have to be experimentally established because for luminescence dosemeters, they depend not only on the effective atomic number (ratio of mass energy absorption coefficients of dosemeter and tissue) of the detector, but also considerably on the luminescence efficiency and the material surrounding the dosemeters. Metal filters are generally used for the compensation of energy dependence below 200 keV and/or to provide photon energy discrimination. It is noted that the contribution to Hp(0.07) could be measured more accurately than Hp(10). For the dosemeters exhibiting high photon energy-dependent response, estimation of the beta component of Hp(0.07) becomes very difficult in the mixed field of beta radiation and photons of energy less than 100 keV. Recent studies have shown that the thickness and the atomic number of metal filters not only affect the response below 200 keV but also cause a significant over-response for high energy (>6 MeV) photons often encountered in the environments of pressurised heavy water reactors and accelerators.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call