Abstract

A method that uses an integrating sphere as a platform for photoacoustic measurement of the Grüneisen parameter Γ of absorbing liquids is developed. Derivation of a simple equation for determining Γ is presented. This equation only requires the voltage peak-to-peak value of the photoacoustic signal detected by a flat transducer and the relative energy of the incident light measured by a photodetector. Absolute detector sensitivities are not required. However, a calibration procedure is necessary. An experimental setup is constructed in order to implement and verify the method. Aqueous ink solutions are used as absorbing liquids to determine the calibration (instrument) constants. Validation of the equation is done by determining Γ of ethanol at room temperature. The obtained value of Γ(ethanol) = 0.72 ± 0.06 has a 7% relative difference to the calculated value from known thermal properties reported in literature.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.