Abstract

Fully inverting spins, instead of merely saturating them, provides superior contrast for tagging procedures. The resulting improvement in tag contrast-to-noise ratio (CNR) yields higher-precision tag detection. Also, thinner slices and hence reduced tag separations can be employed, providing displacement and strain measurements with better spatial resolution. Alternatively, the improved tag contrast can be used to obtain cine images covering a greater portion of the cardiac cycle. The use of standard magnitude reconstruction for images of these inversion tags causes rectification of the negative-valued signals from the tags, confounding the image interpretation. Therefore, a phase-sensitive reconstruction scheme of the inverted tags must be employed. Here we demonstrate the implementation of inverted tags with phase-sensitive reconstruction in a ramped-flip-angle, steady-state free precession (SSFP) sequence.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.