Phase instability is a serious problem in swept-source optical coherence tomography (OCT) with polygon tunable lasers; however, these devices have additional issues. We found that polygon tunable lasers also have fluctuations in output power and sweep range: the former creates artifacts that may impair the recognition of sample information, and the latter reduces the interference signal utilization during phase correction. We demonstrate a method that uses the calibration signal to quantify these problems and improve system stability and image quality. The proposed amplitude correction and phase correction methods are used to eliminate vertical artifacts and improve the resolution of OCT flow and intensity images while reducing the phase error.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.