Abstract
The construction of the adapted patient equivalent phantom (APEP) to simulate the X-ray scattering and absorption by chest, skull and pelvis of nonstandard patient in conventional radiographic equipment is presented. This APEP system is associated to the pre-existing realistic-analytic phantom (RAP) [Pina, D.R., Duarte, S.B., Ghilardi Netto, T., Trad, C. S., Brochi, M.A.C., Oliveira, S.C. de, 2004. Optimization of standard patient radiographic images for chest, skull and pelvis exams in conventional X-ray equipment. Phys. Med. Biol. 49, N215–N226] forming the coupled phantom (RAP–APEP), which is used to establish an optimization process of radiographic images of chest, skull and pelvis for nonstandard patients. A chart of the optimized radiographic technique is established covering a wide range of nonstandard patient thickness, and offering a dose reduction in comparison with those techniques currently used. Different validation processes were applied to confirm the improving of the radiographic image quality when techniques of the established chart are used.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.