Abstract
A method for the determination of the SAR distribution produced by an electromagnetic applicator for localized hyperthermia is described. The procedure for SAR evaluation consists of recording the time evolution of the temperature inside a polyacrylamide phantom by means of thermochromic liquid crystal sheets inserted in it. The technique allows a complete characterization of applicators in a very broad frequency range, using power levels of the order of those needed in real treatments. Criteria for the minimum phantom size and the maximum time duration of the characterization procedure are indicated, which allow a reliable determination of the effective field size and penetration depth of the applicator.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.