Abstract
Almost all high spatial resolution positron emission tomography (PET) detectors based on pixelated scintillator arrays utilize crystal arrays with smaller pitches than photodetector arrays, leading to challenges in resolving edge crystals. To address this issue, this paper introduces a novel multi-resolution silicon photomultiplier (SiPM) array design aimed at decreasing the number of readout channels required while maintaining the crystal resolvability of the detector, especially for edge crystals. The performance of a pseudo 9 × 9 multi-resolution SiPM array, consisting of 6.47 × 6.47 mm2, 6.47 × 3.07 mm2, and 3.07 × 3.07 mm2 SiPMs, was compared to those of a pseudo 8 × 8 SiPM array with a 6.8 mm pitch, and a 16 × 16 SiPM array with a 3.4 mm pitch using a 36 × 36 LYSO array with a pitch of 1.5 mm. The large-size pseudo SiPMs were implemented by digitally grouping multiple 3.07 × 3.07 mm2 SiPMs. The flood histograms show that the edge crystal resolvability of the pseudo 9 × 9 multi-resolution SiPM array is comparable to that of the 16 × 16 SiPM array and is significantly better than that of the 8 × 8 SiPM array.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE transactions on radiation and plasma medical sciences
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.