Abstract

Complex personalized Functional Electrical Stimulation (FES) protocols for calibrating parameters and electrode positioning have been proposed, most being time-consuming or technically cumbersome for clinical settings. Therefore, there is a need for new personalized FES protocols that generate comfortable, functional hand movements, while being feasible for clinical translation. To develop a personalized FES protocol, comprising electrode placement and parameter selection, to generate hand opening (HO), power grasp (PW) and precision grip (PG) movements, and compare in a pilot feasibility study its performance to a non-personalized protocol based on standard FES guidelines. Two FES protocols, one personalized (P1) and one non-personalized (P2), were used to produce hand movements in twenty-three healthy participants. FES-induced movements were assessed with a new scoring scale which comprises items for selectivity, functionality, and comfort. Higher FES-HSS scores were obtained with P1 for all movements: HO (p= 0.00013), PW (p= 0.00007), PG (p= 0.00460). Electrode placement time was significantly shorter for P2 (p= 0.00003). Comfort scores were similar for both protocols. The personalized protocol for electrode placement and parameter selection enabled functional FES-induced hand movements and presented advantages over a non-personalized protocol. This protocol warrants further investigation to confirm its suitability for developing upper-limb rehabilitation interventions with clinical translational potential.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call