Abstract

Gadolinium diethylenetriamine penta-acetic acid (Gd-DTPA) was injected intravenously for magnetic resonance imaging in patients with impaired renal function, and the permeabilities of two types of general hemodialysis (HD) membrane to this agent were compared. One of the HD membranes, Nipro FB-210-U membrane (U membrane), has a larger pore size than the other, Nipro FB-210-M membrane (M membrane). On the same day as magnetic resonance imaging was performed, the patient underwent a 4-hour HD session. Blood was sampled before and at five points after the start of HD. Mean dialysance of Gd-DTPA was 175.8 to 185.8 mL/min with the U membrane and 120.5 to 130.1 mL/min with the M membrane. The elimination rate was 91.1% with the U membrane and 81.1% with the M membrane 4 hours after the start of HD. The permeability was significantly higher with the U membrane than with the M membrane. Although it is more expensive, the U membrane should be chosen rather than the M membrane for HD to remove Gd-DTPA from the body as efficiently as possible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.