Abstract

The performance of the IEA dose equivalent meter based on the REM-2 recombination chamber was tested in pulsed high energy radiation field at CERN-EC calibration facility. The device was working with its own monitoring circuit, and provided accurate and stable results, within 2% of statistical uncertainty.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call